Справочник Пользователя для Rigol MSO1074Z-S 4-channel oscilloscope, Digital Storage oscilloscope, MSO1074Z-S
![Rigol](https://files.manualsbrain.com/attachments/4ef153c5b3a9e56ac40ea738ae281ea9dc808cdc/common/fit/150/50/a84f25c70c016e6b8d8e8d0fbe042c3bbe7b8a8feeee0ecbbc25ec79cb57/brand_logo.gif)
Модели
MSO1074Z-S
RIGOL
2-46
DS1000Z Programming Guide
:MASK:SOURce
Syntax
:MASK:SOURce <source>
:MASK:SOURce?
:MASK:SOURce?
Description
Set or query the source of the pass/fail test.
Parameter
Name
Type
Range
Default
<source> Discrete
{CHANnel1|CHANnel2|CHANnel3|CHANnel4}
CHANnel1
Explanation
This command can only select the channel enabled. You can sent
the
Return
Format
The query returns CHAN1, CHAN2, CHAN3 or CHAN4.
Example
:MASK:SOURce CHANnel2
Related
Command
:MASK:OPERate
Syntax
:MASK:OPERate <oper>
:MASK:OPERate?
:MASK:OPERate?
Description
Run or stop the pass/fail test, or query the status of the pass/fail test.
Parameter
Name
Type
Range
Default
<oper>
Discrete
{RUN|STOP}
STOP
Explanation
the pass/fail test.
Return
Format
The query returns RUN or STOP.
Example
:MASK:OPERate RUN
Related
Command