Справочник Пользователя для Rigol MSO1074Z-S 4-channel oscilloscope, Digital Storage oscilloscope, MSO1074Z-S

Модели
MSO1074Z-S
Скачать
Страница из 172
RIGOL 
DS1000Z Programming Guide 
2-79
 
:TIMebase Commands 
 
:TIMebase commands are used to set horizontal parameters, such as enabling the delayed sweep, setting 
the horizontal timebase mode, etc. 
 
Command List: 
 
 
:TIMebase:DELay:ENABle 
Syntax 
:TIMebase:DELay:ENABle <bool> 
:TIMebase:DELay:ENABle? 
Description 
Enable or disable the delayed sweep, or query the status of the delayed sweep.   
Parameter 
Name 
Type 
Range 
Default 
<bool> 
Bool 
{{0|OFF}|{1|ON}} 
0|OFF 
 
Explanation 
Delayed sweep can be used to enlarge a length of waveform horizontally to view the 
waveform details.
 
Return Format 
The query returns 0 or 1.   
Example 
:TIMebase:DELay:ENABle ON 
 
 
:TIMebase:DELay:OFFSet 
Syntax 
:TIMebase:DELay:OFFSet <offset> 
:TIMebase:DELay:OFFSet? 
Description 
Set or query the delayed timebase offset. The default unit is s. 
 
Parameter 
Name 
Type 
Range 
Default 
<offset>  Real 
-(LeftTime  -DelayRange/2) to   
(RightTime  -DelayRange/2) 
 
Explanation 
LeftTime = 6×MainScale – MainOffset 
RightTime = 6×MainScale + MainOffset 
DelayRange = 12×DelayScale
 
Return Format 
The query returns the offset in scientific notation.   
Example 
:TIMebase:DELay:OFFSet 0.000002   /*Set the delayed timebase offset to 2μs*/ 
:TIMebase:DELay:OFFSet?   /*The query returns 2.0000000e-06*/ 
Related 
Command