Справочник Пользователя для Benning IT 120 B 044102

Модели
044102
Скачать
Страница из 80
18
Resistance of auxiliary earth electrode and probe too 
high.
4.1.4  Result field
Measurement passed.
Measurement failed.
Measurement is aborted. Check the conditions at the in-
put terminal.
4.1.5  Other messages
Instrument settings and measurement parameters/limits 
are set to initial (factory) values. For more information re-
fer to chapter 4.5.4 Recalling original settings.
Probe
LUXmeter probe is turned off or disconnected from the 
instrument.
First measurement
The first stored measurement result is displayed.
Last measurement
The last stored measurement result is displayed.
Memory full
All memory locations are occupied.
Already saved
Measurement results already saved.
CHECK SUM ERROR
Important memory contents lost or damaged. Contact 
your distributor or manufacturer for further information.
4.1.6  Sound warnings
Shortest sound
Pressed key deactivated. 
Sub-function is not available.
Short sound
Pressed key activated.
Measurement has been started after pressing the 
TEST key. Consider any displayed warnings during 
measurement.
Long sound
Measurement prohibited. Consider any displayed war-
nings and check online voltage/terminal monitor.
Periodic sound
Warning! Phase voltage on the PE terminal! Stop all the 
measurements immediately and eliminate the fault before 
proceeding with any activity.