Note:
-  It is useful to deactivate the COMMANDER, if strong sources of interference might affect the 
functioning of the COMMANDER.
4.4.6  Recalling original settings
The following parameters and settings can be set to initial (factory) values:
-   Test parameters and limit values 
-   Contrast
-   Prospective short/fault current scaling factor
-   Supply system
-   Communication port (BENNING IT 120 B only)
In order to recall the original setting press and hold the 
 key and switch on the instrument. “Hard 
reset” message will be displayed for a while.
Instrument settings, measurement parameters and limits are set to their initial values as follows: 
Instrument settings
Default value
Contrast
50 %
Prospective short/fault current scaling 
factor
1.00
Supply system
TN/TT
Communication port
RS 232
Function  
Sub-function
Parameter / limit value
CONTINUITY
Sub-function: RLOW
LowΩ resistance
High limit resistance value: 2.0 Ω
Continuity
High limit resistance value: 20.0 Ω
INSULATION RESISTANCE
Nominal test voltage: 500 V
Low limit resistance value: 1 MΩ
IMPEDANCE
(Zl)
(Zs)
Fuse type: none selected (F)
Fuse current rating: none selected (A)
Fuse trip-out current: none selected (ms)
RCD
Sub-function: RCD Uc
Contact voltage – RCD Uc
Trip-out time – RCD t
Trip-out current – RCD III
Autotest – RCD AUTO
Nominal differential current: I
∆N
=30 mA
RCD type and test current starting polarity:<
 G 
Limit contact voltage: 50 V
Nominal differential current multiplier: 
×1
LOOP IMPEDANCE
Testing RCDs with nominal differential current ≥ 10 mA
RESISTANCE TO EARTH
(BENNING IT 120 B only)
High limit resistance value: 50 Ω
ILLUMINATION  
(BENNING IT 120 B only)
Low limit illumination value: 300 lux