Auto Mode Testing
Background
The Auto Mode function allows you to run two
test functions back to back automatically. Only
after both tests have run will a PASS/FAIL
judgment be determined on the tests.
test functions back to back automatically. Only
after both tests have run will a PASS/FAIL
judgment be determined on the tests.
Steps
1. Connect the DUT to the tester.
2. Ensure the tester is in READY
status and that all the testing
parameters have been set for both the
tests* that you wish to run in Auto
Mode.
tests* that you wish to run in Auto
Mode.
From Page
33
33
3. Press the Mode key to toggle
between each of the Auto Modes.
O D E
M
Each mode indicates the order that the tests will
execute in. For example:
IR-DC indicates that an IR test will be followed
by a DCW test.
execute in. For example:
IR-DC indicates that an IR test will be followed
by a DCW test.
Range
MODE, AC-IR, IR-AC, DC-IR, IR-
DC
Note: MODE is the default setting
which only runs a single test a time.
DC
Note: MODE is the default setting
which only runs a single test a time.
H = 2 2 . 0 m A
R
m A
L
0
=
0 0
.
7 4 S
= 1
T
CW
D
A CW
R
I
-
C
I
A
T M E
I
Selected Auto mode
1 00
k V
A
E
R
D
Y
A
m
4. The Auto Mode function is now ready to test