Gw Instek GPT-9601 Insulation measuring device, GPT-9601 Benutzerhandbuch

Produktcode
GPT-9601
Seite von 75
 
OPERATION 
49 
Auto Mode Testing 
 
Background 
The Auto Mode function allows you to run two 
test functions back to back automatically. Only 
after both tests have run will a PASS/FAIL 
judgment be determined on the tests. 
Steps 
1.  Connect the DUT to the tester. 
2.  Ensure the tester is in READY 
status and that all the testing 
parameters have been set for both the 
tests*
 that you wish to run in Auto 
Mode.  
From Page 
33 
 
3.  Press the Mode key to toggle 
between each of the Auto Modes. 
O D E
M
 
 
 
Each mode indicates the order that the tests will 
execute in. For example: 
IR-DC indicates that an IR test will  be followed 
by a DCW test. 
 
 
Range 
MODE, AC-IR, IR-AC, DC-IR, IR-
DC 
Note: MODE is the default setting 
which only runs a single test a time. 
 
 
     
 
 
 
H = 2 2 . 0 m A  
 
 
 
m A
L
0
=
0 0
.
   
7 4 S  
 
 
 
 
 
= 1
T
       
   
CW
D
A CW
R  
I
-
C
I
A
T M E
I
Selected Auto mode
 
1 00
 
k V
A
E
R
 
D
 
Y
A
m
 
 
4.  The Auto Mode function is now ready to test 
two tests in succession.