Справочник Пользователя для Beha Amprobe GT-800 PRO KITVDE-tester 4472081

Модели
4472081
Скачать
Страница из 76
GT-600/GT-800 Instruction Manual 
20 
 
 
 
Pass/fail limit value calculation: 
 
Cross section A 
Cable length L 
Limit value 
 1.5 mm
2
 
 5 m 
0.3 
 
 5 ... 12.5 m 
0.4 
 
 12.5 ... 20 m 
0.5 
 
 20 ... 27.5 m 
0.6 
 
 27.5 ... 35 m 
0.7 
 
 35 ... 42.5 m 
0.8 
 
 42.5 ... 50 m 
0.9 
 
 50 m 
1.0 
 
2.5 ... 50 mm
2
 
1 ... 50 m (res. 0.5 m)
R
LIM
 = 
.L/A 
 
Where: 
 ...  0.01786 mm
2
/m (specific resistance of Cu) 
L  ..  cable length in m 
A ..  cross section in mm
2
 
 
Specific information that can be shown on the display: 
 
Information displayed 
Description 
TEST LEAD 
NOT COMPENSATED 
Test lead is not compensated (no “C” mark on 
the display)! It is recommended to 
compensate the test lead. 
>5
, NOT COMPENSATED 
Compensation was not successfully done due 
to too high compensation value (>5.00 
)! 
COMP? 
Test lead is wrong compensated (negative R
PE
 
result 
0.05 ). Compensate the test lead 
again. 
 
 EXTERNAL VOLTAGE 
External voltage is applied to one or more 
test terminals, see the explanation in section 
“Pre-Tests and Protection, Functions and 
Messages”. 
 
 USE PROBE 1 
Test lead is connected to PROBE 2. Reconnect 
it to PROBE 1. 
>11.00
 
X
 
R
PE
 value higher than 11.00 
 (over-range), 
possibly because of open test lead. 
 
  In order to save displayed measurement result, press the “SAVE” function 
key (F5) twice, for further instructions see the “Memorizing example” 
section.