Manual test procedure means that you have to confirm every test
stage by pressing the [OK
] button to get to the next test stage.
Automatic test procedure means that a test stage automatically
changes to the next after 5 seconds - except for confirmation of
connection to the mains voltage (see Point 6.6, basic settings of
the tester)
The [Cancel] button takes you back to the ‘Main menu’.
To move to the next stage - also see Diagram 2 in Point 9.9.
12.1 
 Visual inspection
The [->Next] button takes you to the first stage of the test, the visual
inspection (Fig. 54).
The housing, connecting cable, inscriptions and other parts should be
inspected in this test.
You confirm positive results of the visual inspection by pressing the
appropriate buttons – the display changes from ‘not OK’ to ‘OK’ (Fig. 55).
12.2 
Insulation resistance
The limit value is:
0.25 M
Ω (PC III)
After pressing the [->Next] button, the display for the insulation resistance
measurement appears (Fig. 56).
Beside the measured value, the level of the test voltage is also displayed
(min. 500 V DC).
The star sign on the left side of the display blinks when the
measuring process is running (the values are being measured).
The [OK] button appears when the measured value is determined.
A help display with a measuring principle circuit and a help text is also
available here.
Since the tester cannot provide the specific low voltage for the test
object, a function test is not carried out for test objects in PC III.
Alternatively, the test object can be tested according to PC II with
the appropriate power supply unit - see Point 11 on Page 29.
36
Operating manual for the TG uni 1 appliance tester
Conducting tests in accordance with VDE 0701 and 0702 - PC III
Fig. 54
Fig. 55
Fig. 56