Справочник Пользователя для SSI America BINARY SOLVENT DELIVERY MODULE 90-2581 REV B

Скачать
Страница из 48
 
 A-4 
PI OK,a.aa,b,c,d,e,f,g,h,i,j,k,l,
m,n,o,p,q/ 
 
(a.aaa, a.aa, aa.aa, or 
aa.a) 
(c or cc) 
Reads the current pump setup, where: 
 
a.aaa, a.aa, aa.aa, or aa.a = Flow rate in mL/min 
 
b = Run status (0 = stopped, 1 = running) 
 
c or cc = Pressure compensation 
 
d = Pump head type (see RH command) 
 
e = Pressure Board present = 0; otherwise 1 
 
f = External control mode (0 = frequency, 1 = 
voltage) 
 
g = 1 if pump started and frequency controlled, 
else 0 
 
h = 1 if pump started and voltage controlled, else 0
 
i = Upper pressure limit fault (0 = no, 1 = yes) 
 
j = Lower pressure limit fault (0 = no, 1 = yes) 
 
k = Priming (0 = no, 1 = yes) 
 
l = Keypad lockout (0 = no, 1 = yes) 
 
m = PUMP-RUN input (0 = inactive, 1 = active) 
 
n = PUMP-STOP input (0 = inactive, 1 = active) 
 
o = ENABLE IN input(0 = inactive, 1 =active) 
 
p = Always 0 
 
q = Motor stall fault (0 = no, 1 = yes) 
RE 
OK/ 
Resets the pump configuration to its default power-up 
state. 
 
A.2 
Pump Rear Panel 4-Pin and 10-Pin Terminal Board Connectors  
A 4-pin terminal board connector and a 10-pin terminal board 
connector are provided on the back panel.  Any device capable of 
providing the proper run/stop logic level, flowrate control frequency, 
or flowrate control voltage can be used as a remote controlling device 
for pump operation via this connection.  The terminal board 
connectors can be removed for ease of connecting wires, if desired, by 
pulling firmly rearward and should be reinserted firmly afterward. 
A.2.1  Pressure Fault and Motor Stall Fault Output 
The pump's output is on the 4-pin terminal board connector.  The 
pinout is: 
 
 
Pin 
Function
 4 
EVENT 
 3 
EVENT 
 2 
EVENT 
 1 
Ground 
 
This output is produced internally by a reed relay which has SPDT 
contacts with a 0.25 amp maximum, 50 VDC maximum, 0.2 ohm 
rating.  The 4-pin connector allows wires to be connected to the 
EVENT 1 (Pole), EVENT 2 (NC), and EVENT 3 (NO) terminals.  
When the pump stops due to the sensed pressure exceeding the set 
pressure limits or if a motor stall fault occurs, the connection between