FP-AI-112 and cFP-AI-112
Caution
Cascading power between two modules defeats 
isolation between those modules. Cascading power from 
the network module defeats all isolation between 
modules in the FieldPoint bank.
Measuring Voltage with the [c]FP-AI-112
Figure 3 shows how to connect a voltage source without an 
external power supply to one channel of the [c]FP-AI-112.
 
Figure 3.  Voltage Source without External Power Supply
Figure 4 shows how to connect a voltage source with an external 
power supply to one channel of the [c]FP-AI-112.
 
Figure 4.  Voltage Source with External Power Supply
V
SUP
V
IN
V
C
>100 MΩ
Input
Impedance
COM
[c]FP-AI-112
Voltage
Source
+
Input 
Circuitry
V
SUP
V
IN
V
C
>100 MΩ
Input
Impedance
COM
[c]FP-AI-112
Input 
Circuitry
Powered
Voltage
Transducer
OUT
+
+
External 
Power Supply
(Optional)
2 A
Max