SAFETY PROTECTIONS
The nebulizer is equipped with protection against 
several unsafe situations. If  these situations occur, the 
nebulizer will either (1) not power up or (2) turn off  
automatically, as described below.
Protection against incorrect Power Supply
If  an AC/DC adapter is used to power the device, 
ensure the adapter meets local power regulations; if  
not, the device may not power up (the Power Indicator 
Light will not illuminate). 
Protection Against Overheating
During normal operation, the Power Indicator Light 
will illuminate GREEN. However, in the two following 
cases, the device will indicate irregularity by turning 
off  automatically. The device then requires approxi-
mately 45 minutes at rest before restarting.
Protection Against Overuse: The electronic compo-
nents of  the device are designed to function continu-
ously for approximately 40 minutes. Operating the 
device for too long will trigger the heat-protection 
mechanism and the device will turn off  automatically.
Lack of Liquid in the Reservoir: In the event of  lack or 
insufficient liquid in the reservoir, the device will turn 
off  automatically. 
6700 TECHNICAL SPECIFICATIONS
Power Supply: AC/DC Adapter (240V, 50Hz; 
12VDC 800mA)
Power Consumption: ~ 12W (12V/1A)
Sound Level: < 40dB
Frequency: 2.8 Mhz
Particle Range: 0.5~4.0 microns
Medication Cup Capacity: 5 ml maximum
Average Rate: 0.7 ml/min. maximum 
(NaCl0.9% Normal Saline)
Intermittent Use: 45 min. on/ 45 min. off
Operation Conditions: 50°F ~ 104°F 
(10°C ~ 40°C), RH 10% ~ 95%
Storage Conditions: -13°F ~ 140°F 
(25°C ~ 60°C), RH 10% ~ 95%
Dimensions: ~ 5.1 x 3.3 x 2.2 in. 
(~ 130 x 85 x 55 mm)
Weight without batteries: ~ .44 lb (198 g)