  For Safe Handling 
SAFETY - 4
 
 
  Warnings in the Operating Manual 
 
Warnings provided in this manual and where they occur are summarized below. 
 
 
●  This system does not have components that require 
  Danger 
signs. 
 
  Warning Indicators 
 
 
●  Burns from the Rotary Pump 
 
■  If it stops abnormally the rotary pump remains hot.    Touching it 
can cause burns.    Do not touch the rotary pump in the event of 
an abnormal stoppage. 
(Chapter 4 Maintenance) 
 
●  Burns from Touching the Filament Assembly 
 
■  When the filament assembly is replaced, it can be as hot at 90℃.  
Touching it can cause burns.    When replacing a filament 
assembly, turn off the acceleration voltage, and let the assembly 
stand for 30 minutes or longer after the air is let in before working 
on it. 
Or, execute the filament exchange by using heatproof gloves of 
the clean room specification. 
(Chapter 4 Maintenance) 
 
●  Beware of Electric Shock 
 
■  Voltages up to 100 V AC and 30 kV DC are used inside this 
instrument.    Do not touch inside or you may receive an electric 
shock. 
Do not remove the covers of main unit and display unit and do 
not touch internal parts or circuits while the instrument power is 
turned on.    There is a hazard of fatal or serious injury due to 
electric shock. 
(Chapter 4 Maintenance)