Инструкции Пользователя для Bacharach Informant 2

Скачать
Страница из 32
The Informant™ 2 
 
 
20 
 
0019-9211 Rev 9 
Symptom 
Probably Cause and Remedy 
Short refrigerant 
sensor life. 
 
a. Probe filter dirty, thus not allowing 
enough air to flow over the sensor to 
keep it cool. Replace filter often per 
Section Probe Filter Replacement. 
b. Sensor has often been exposed to high 
concentrations of refrigerant. Avoid 
sampling pure refrigerant for long 
periods of time. 
Slow gas response 
time. 
 
a. Probe filter clogged. Replace filter per 
Section Probe Filter Replacement.  
b. Sensor nearly depleted or faulty. 
Replace sensor per Section Sensor 
Replacement
.  
c. Crack in rubber sheathing that covers 
the probe’s flexible tube, allowing air to 
enter through the crack. Replace probe 
assembly per Section Probe Assembly 
or Fan Replacement
d. Internal fan defective. Replace fan per 
Section Probe Assembly or Fan 
Replacement.
 
Erratic gas indication. 
 
a. Sensor faulty. Replace sensor per 
Section Sensor Replacement.  
b. Loose sensor. Unscrew probe tip and 
ensure that sensor is being held firmly 
in socket. 
c.  Loose printed circuit board connector. 
Disassemble detector and ensure that 
all connectors are securely attached to 
the printed circuit board. See Figure 8.