Справочник Пользователя для IBM ECS-320A

Скачать
Страница из 63
S
ERIAL 
I
NTERFACE 
D
EVELOPERS 
G
UIDE
 
 
Bits 10-11 (Pfv Src Sel): Processed FPA Video Source Select: 0 – Digital FPA Video; 1 – NUC 
Corrected Video; 2 – Pixel Replaced Video; 3 – ITT Video 
6.9.4 ATC Offset Coefficient Register  
Address: 0x4007 
16-Bit Signed Constant Added to Output of NUC Circuit (Prior to NUC Refresh) 
-16,384 <= Atc Ofst <= 16,383.5 
6.10 Access to Serial Data Flash 
The Atmel serial data flash chip has 4096 pages of storage each with 264 bytes.  The pages have 
been allocated as follows for the embedded application. 
U
Page(s)
U
 
U
Allocated Use 
0 - 7 
Reserved 
8 - 23 
Op Mode Descriptor Tables (16x) 
24 - 87 
NUC Mode Descriptor Tables (64x) 
88 - 127 
Reserved 
128 - 159 
Zone Statistic Descriptor Tables (32x) (Currently not used) 
160 - 190 
Reserved 
191 
Overlay Palettes (8x) 
192 - 223 
Color Palette Y Tables (16x) 
224 - 255 
Color Palette Cr/Cb Tables (16x) 
256 - 319 
User Defect Pixel Lists (64x) (Currently not used) 
320 - 511 
Reserved 
512 - 1145 
FPA Processor FPGA Configuration File 
1146 - 1535 
Expansion FPGA Configuration File 
1536 - 2175 
Stored Image 0 (Currently not used) 
2176 - 2815 
Stored Image 1 (Currently not used) 
2816 - 3455 
Stored Image 2 (Currently not used) 
3456 - 4095 
Stored Image 3 (Currently not used) 
The host can read/write to these locations through the use of the CMD_COPY_SFLASH_PAGE, 
CMD_PROG_SFLASH_FULL, CMD_PROG_SFLASH_PARTIAL commands.  These commands are 
implemented in Lumitron’s host applications to load palettes, FPGA configuration files, operational 
mode descriptors, and NUC mode descriptors. 
7 Remote Calibration Process 
The process to complete a non-uniformity calibration is initiated with a Lumitron defined command. 
After the calibration command has been issued, the embedded application communicates with the 
host based on the state of the “CameraConfig.camStats.HostStatusCode” global structure member.  
This variable, hereafter referred to as status code, has its state set by either the embedded application 
or the host application depending upon the next step in the process. 
Status codes that are used in the calibration or defective pixel detection process are as follows: 
38