UVPS 81-0112-01 Rev C 用户手册

下载
页码 22
UV CROSSLINKERS 
 
UVP 
 
 
6.  To secure the starter holder, place the holder in your right hand and with your left hand in 
the chamber, press the now loosened reflector adjacent to the holder inwards. This will 
push the black spigots inwards thereby securing the starter holder. 
 
7.  Re-fit the reflector. 
 
Replacement of fuse holders 
 
These are located at the back of the unit just above the mains connector. There are two fuses in 
separate holders. Replacement is the same for both. 
 
1.  Cut back the heat shrink on the two wires. 
 
2.  Unsolder both connections and pull from the holder. 
 
3.  Unscrew the nylon nut and push the holder out through the back of the unit. 
 
4.  To replace-reverse the above procedure. 
 
Replacement of main connector 
 
This is located at the back of the unit below the fuse holders. 
 
1.  Cut back the heat shrink on the three wires. 
 
2.  Unsolder the three connections and pull from connector. 
 
3.  Release main connector by holding the nuts with a spanner and then on the outside back of 
the unit, release the two screws one at a time (using a screwdriver). The main connector 
should now be loose. 
 
4.  To replace-reverse the above procedure. 
 
Replacement of main switch 
 
This is located on the front panel below the membrane pad. 
 
1.  Remove the wires noting their position. 
 
2.  Remove the switch by pushing it out from the back towards the rear of the front panel. 
 
3.  Push in the new switch and re-connect. 
 
 
 
- 15 -