11. MAINTENANCE 
 
11-1. Notes on maintenance 
‰
 Do not disassemble the balance. Contact your local A&D dealer if your 
balance needs service or repair. 
‰
 Please use the original package for transportation.   
‰
 Do not use organic solvents to clean the balance. Use a warm lint free cloth 
dampened with a mild detergent. 
‰
 To clean around the pan support, refer to the drawings below. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
EJ-120 / EJ-200 / EJ-300 
EJ-1500 / EJ-2000 / EJ-3000 
EJ-410 / EJ-610 
EJ-4100 / EJ-6100 
 
 
11-2. Error codes 
Overload error 
Warning to indicate that an object beyond the balance 
capacity has been placed on the pan. Remove the object 
from the pan. 
 
Range over notice 
This will be shown if the weight sensor receives strong 
force upward. Check if there is anything sandwiched 
around the weighing pan. There is a possibility that the 
weight sensor itself may have a failure. 
 
Unit weight error 
The sample weight is too light to set the unit weight in the 
counting mode or 100% reference weight in the % mode. 
 
 
Sample quantity notice 
When sample weight is light and the counting error could 
become large, the balance will request you to use larger 
number of samples. Place the displayed number of 
samples on the pan and press the    PRINT    key to store 
 
 