B&K 8510 用户手册

下载
页码 57
 
16
GROUP 
Total = 1000 steps 
1 1000 
steps 
500
00
00
00
 steps 
500 steps 
250 steps 
250 steps 
250 steps 
250 steps 
120 steps  120 steps  120 steps  120 steps  120 steps  120steps  120 steps  120 steps
 
When receiving one trigger signal, it will start the list operation until receiving another trigger signal or finish the List 
operation. 
 
Trigger
1
0
2
3 4
5
List count=1
List count=2
List sequence
 
Triggered Operation 
The electronic load has various triggering modes to allow synchronization with other test equipment or events. Such as: 
Keypad triggering mode: Press 
Shift
+
Trigger
  to trigger the electronic load.   
TTL triggering mode: Send a high pulse with a constant time more than 5m Sec to the trigger terminals in rear panel 
to trigger the electronic load.   
Command triggering mode: Send triggering command to the electronic load via the serial port. 
 
Input Control 
Short On/Off 
The DC Load can simulate a short circuit at its input by turning the load on with full-scale current. The short circuit can 
be toggled on/off at the front panel using the 
Shift
+
Short
  keysShort
 operation is not influence the 
operation setting current value , When short operation is on OFF state , the Load reverts back to the original setting 
state.   
The actual value of the electronic short is dependent on the mode and current range that are active when the short 
is turned on. 
In CC, CW and CR mode, the max short-circuit current value is 1.2 times of the current range. In CV mode, short-circuit 
operation is same as the operation of setting CV to 0V.   
     
NOTE
 
Turning the Short Test on in CV mode may cause the load to draw so much current that the software current 
limit is triggered, which may turn the input off. Turning the short circuit on does not affect the programmed 
settings, and the load input will return to the previously programmed values when the short is turned off.