Trigger and Remote Sensing Connections
 
A 4-pin connector and a quick-disconnect mating plug are provided on rear panel for accessing input signals and 
remote sensing, all leads connected to the connector should be twisted and shielded to maintain the instrument's 
specified performance. 
 
Remote Sensing: 
sense ( +
++
+ )
 )
 )
 )    and    sense    ((((    ----    )))) Used to connect the remote sensing leads to the power source.
 
 
TRIG IN: TRIG (IN) 
A TTL-compatible input that responds to external edge trigger signal. A trigger applied to this 
input can be used to change settings (voltage, current, resistance, etc.), toggle between settings in transient-toggle 
mode, or generate a pulse in transient-pulse mode.
 
 
TRIG GND: TRIG (
) 
Provides the common connection for the trigger signals. 
 
Turn-On Checkout 
Introduction 
Successful tests in this chapter provide a high degree of confidence that the electronic load is operating properly. 
Checkout Procedure 
The test in this section checks for proper operation of the electronic load. If you have not already done so, connect the 
power cord to the unit and plug it in
Procedure 
Display 
  Explanation 
1. Turn the unit on.        The 
electronic Load undergoes a 
self-test when you First turn it 
on. 
 
During self test, all segments are briefly 
lit 
 
2. Wait for 1s after turn on 
electronic load. 
EPROM ERROR 
EEPROM damage or Lost data of last 
power off 
Run well if no such display, system will 
go to the step 3 directly. 
3. Wait for another 2S.     
ERROR CAL.DATA 
EEPROM Lost calibration data 
Run well if no such display, system will 
go to the step 4 directly. 
4. Press Shift button and  △▽
keys . 
LOAD MODEL:85XX 
SN:XXX-XXX-XXX 
VER x.xx 
Display the information of the product 
Type, series number version of 
software. 
5. Press 
ESC
 button   
0.000V   0.000A 
Display the actual input voltage and 
current value.