B&K 8510 用户手册

下载
页码 57
 
25
6.Press  △▽   
0.000W   I: 0.000A 
Display the actual power value and 
setting value. 
In Case of Trouble   
Should the power-on self test of the Electronic Load fail, you may try the following:    When turning on the power, make 
sure the power cord to the unit is plugged in and the power switch is in the on position. 
1) 
Check the power voltage setting. 
Work voltage of load have two type 110V or 220V , Please make sure it is right voltage accordance to the voltage 
in your area. You could change the voltage setting through dial the switch in the rear panel.   
 
2) Check 
the 
fuse. 
If fuse is blown, please replace it according to the following specification. 
 
Model  
Fuse specification 
(110VAC) 
Fuse specification 
(220VAC) 
8510 
T0.5A  250V 
T0.3A  250V 
 
 
3) 
Location of Fuse   
 
 
Front Panel Operation 
Display  
16-character fluorescent display for showing measurements and programmed values. 
Front Panel Display 
 
Indicators for operating modes and status conditions: 
OFF
 
power off
 
Trigger
 
Indicates that the electronic load is waiting an 
initiate and trigger to occur.
 
CC
 
The selected input channel is in the 
constant current (CC) mode. 
Sense
 
Indicates that the electronic load is in sense 
state
 
CV
 
The selected input channel is in the 
constant voltage (CV) mode. 
Error
 
 
A errors have occurred
 
CW
 
The selected input channel is in the 
constant power (CW) mode. 
Link
 
 
In the communication state
 
 
Fuse