Step 1   
Press
R-set
   
RESISTANCE =0000R 
Step 2   
enter a new value by using numeric keys or Rotary 
knob to adjust the resistance value   
RESISTANCE=*****R 
Step 3   
Press
Enter
 to confirm. 
0.000W     R:0000R 
 
V-set (set up a constant voltage from 1.5V to Max voltage) 
8500A electronic load can be setup for a constant voltage. 
Constant voltage setup procedure is as following: 
Procedure 
Operation details 
VFD display 
Step 1 
Press 
V-set
V-set
 
VOLTAGE=1.500V 
Step 2 
enter a new value by using numeric keys or Rotary 
knob to adjust the resistance value 
VOLTAGE=*.***V 
Step 3 
Press
Enter
 to confirm. 
0.000W    V:3.000V 
Shift + Store   
Procedure 
Operation details 
VFD display 
Step 1 
Press
Shift
and
Store
 
STORE 1 
Step 2 
Press
Enter
 to confirm. 
Store the relative data 
 
Shift + Recall   
Procedure 
Operation details 
VFD display 
Step 1 
Press
Shift
and
Recall
 
RECALL 1 
Step 2 
Press
Enter
 to confirm 
Recall the saving data 
On/Off input setting   
Use
On/Off
to change the state of electronic load. Switch on to off state by press
On/Off
 
Transition Testing Operation 
Users could switch between the two different current and voltage level settings in the transition mode; it could test the 
transition specialty of power supply. Users could use front panel or communication interface(TRAN ON AND TRAN 
OFF)to make it work or not, please setting parameters before transition operation. Include Transition setting value, 
Constant pulse width setting and Transition Pulse width setting and Transition testing mode. The mix pulse width is 
500uS.The Max pulse width is 6S. Transition Operation only could work in CC and CV mode. 
Users can choose one of the three operation modes: Continuous, Pulse and Toggling mode.