STEP 3 
Enter the original value which displayed in 
the VFD or enter a new value by using 
number keys or Rotary knob to adjust the 
voltage value   
CURRENT=*.***A 
STEP 4 
Press 
Enter
 to confirm  
0.000V     *.***A  
 
Setup the input current at 4.33A. 
Method 1: To set up by using number keyboard 
Step1. Press 
I-set
I-set
button. 
Step2. Press numeric button to enter the current value 4.33. 
Step3. Press 
Enter
  button to confirm the current value. 
 
Method 2: To set up
I-set
I-set
  by using Rotary SW 
(1) If the key board is unlocked by password, directly adjust the Rotary SW button, and voltage will be continually 
changed from the previews value according the rotation. At the beginning, the cursor will be shown on the last number 
of the value which is indicated on the VFD, you can move the cursor to the first number, second number etc by using 
number buttons, and then adjust the Rotary SW to change each number, and let it stay at *.** A. Please see the 
following description. Then press I-set to confirm the value. 
0.00A 0.00V   
0.0W 4.33A   
 
Procedure:  
Step1. Press
I-set
I-set
button, 
Step2. Adjust the Rotary knob to change the value, the operation is as the same as item (1) 
Step5. Press
Enter
button to confirm the current value. 
 
P-set (set up a constant power from 0 to Max power) 
8500 series electronic load can be set up for a constant power. 
Constant power setup procedure is as following: 
 
Procedure 
Operation details 
VFD display 
Step 1 
Press
P-set
   
POWER =0.000W 
Step 2 
Enter a new value by using numeric keys or 
Rotary knob to adjust the voltage value   
POWER=*.***W 
Step 3 
Press
Enter
 to confirm. 
0.000W  P:*.000W 
R-set (set up a constant resistance from 0.1Ω to 4000Ω) 
8500 series electronic load can be setup for a constant resistance. 
Constant resistance setup procedure is as following: 
Procedure 
Operation details   
VFD display