B&K 8510 用户手册

下载
页码 57
 
35
Steps: 
1) Press
Shift
+
Menu
to enter into menu, VFD displays
>CONFIG 
2) Press
Enter
into the submenu, VFD displays
>INITIAL CONFIG 
3) Press▼to make VFD display 
RANGE SELECT 
4) Press
Enter
to confirm, and press▼to choose 
ON 
5) Press
Enter
 to confirm 
6) Press
ESC
  to exit menu 
Voltage range: Now you can press 
+▲ to switch from the maximum range to the lower range and high 
resolution.    If the former maximum voltage value was 120V with a resolution of 10mV, it will change to 18V range/ 
1mV resolution when pressing
+▲ , and the display will show 3 decimal points (xx.xxx) as opposed to 2 
decimal points.    If the former maximum voltage was 18V, you can press
+▼ to set the instrument to 120V 
range, with this setting the resolution will change accordingly to 10mV. 
Current range: Press
+▼ to select the maximum current range. If the former maximum current value is 
12A and the resolution is 1mA, it will now be set to 120A max/10mA resolution.    By pressing 
+▼, you 
will get 2 decimal points (xx.xx) compared to 3 decimal points.    To toggle from the 120A to the 12A range, press 
+▲, now the resolution will be set to 1mA. 
 
 
 
Specifications
 
 
Parameter 
8510 
Voltage 0 
 
to 
120V 
Current 1mA 
to120A 
Input rating        
Power 600 
Range 
Accuracy 
Resolution 
0-18V 
±(0.05%+0.02%FS) 
1mV 
0-120V 
±(0.05%+0.025%FS) 
10mV 
0-12A 
±(0.1%+0.1%FS) 
1mA 
 
Load 
Regulation 
0-120A 
±(0.2%+0.15%FS) 
10mA 
0.1-18V ±(0.05%+0.02%FS) 
1mV 
CV Mode 
Regulation 
0.1-120V ±(0.05%+0.025%FS) 
10mV 
0-12A 
±(0.1%+0.1%FS) 
1mA 
CC Mode 
Regulation 
0-120A 
±(0.2%+0.15%FS) 
10mA 
0.1 -10Ω ±(1%+0.3%FS) 
0.001Ω 
10-99Ω ±(1%+0.3%FS) 
0.01Ω 
100-999Ω ±(1%+0.3%FS) 
0. 
1Ω 
CR Mode Regulation 
Input current ≥FS 10% 
Input Voltage≥FS 10% 
1K-4KΩ ±(1%+0.8%FS) 
1Ω