The DB9 interface connector on the rear panel of electronic load is TTL voltage level; you can use the communication 
cable (IT-E132) to connect the DB9 interface connector of the electronic load and the USB interface connector of 
computer for the communication. 
PC
Load 
IT-E131 communication 
cable 
CO
M
P
UT
E
R
IN
S
T
R
U
M
E
N
T
RX
TX
IT-E131  ISOLATED
COMMUNICATION CABLE
TTL(5V)
RS232
ISOLATION
859666668889942311
IT
 
 
 
                     
Note: Do not connect the DB9 connector of the electronic load directly to a PC or other RS232 port. 
 
 
2 Communication between electronic load and PC 
Before using the remote operation mode, please make sure that the baud rate and communication address in 
electronic load are the same as in the computer software, otherwise, the communication will fail, you can change the 
baud rate and communication address from the front panel or from computer. 
1.  Address: the range is from 0 to 254,default setting is 0 
2.  Baud rate: 4800,9600,19200 and 38400 are selectable, default setting is 4800 
3. Data bit:8 bit    
4. Stop bit:1 
5. Parity:None 
 
 
 
3. Frame Format   
 
  Frame length is 26 bytes. Details as following: 
AAH   
Address  Command  4—25bytes are information content 
Parity code 
Description  
 
1.  Start bit is AAH,occupies one byte. 
2.  Address range from 0 to FE, occupies one byte.   
3.  Each command occupies one byte. Following is the command details.   
20H 
Selecting the Remote control mode   
21H 
Selecting the input on/off state   
22H 
Setting the max input voltage   
23H 
Reading the max setup input voltage. 
24H 
Setting max input current   
Stop Bit 
8 Data Bits 
Start Bit
 
Parity=None 
Stop Bit
IT-E132 communication