Toshiba LF470/LF612 规格说明表单

下载
页码 8
 TIC-LF470B 
 
2
 
Model LF470 Detectors 
Fluid pressure: -15 to 150 psi, or -1.0 to 10 bar   
(-0.1 to 1 MPa) 
Principal materials: 
Measurement tube – Alumina ceramics 
Electrodes – Platinum 
Pipe connection port: 
Standard – 316 stainless steel 
Option   Ti (titanium), 
Polyvinyl chloride (shock-resistant)+ 
Ta (Tantalum) for grounding plate, 
Polyvinyl chloride (shock-resistant)+ 
Pt-Ir grounding plate, 
Hastelloy C (Equivalent) 
Dedicated preformed cables: 
Signal cable
 
– 2-wire shielded chloroprene 
sheathed cable 
                      Overall length: 0.28 inch (7mm) 
           Length: 0.20 inch (5m) 
Excitation cable  3-wiren chloroprene sheathed 
cable 
Overall length: 0.28 inch (7mm) 
  Length: 
0.20 inch (5m)
 
Coating:  phthalic acid resin coating pearl-gray 
colored 
Weight: approximately 2.2 lb (1.0 kg)   
                (for each size excluding cables) 
 
 
Model LF612 converters 
Input signals 
Analog signal  the voltage signal from detector, 
proportional to process flow rate (for LF612 
separate type converter). 
Digital input DI (opt.) 
Signal type:    20 to 30Vdc voltage signal 
Input resistance: 2.7k
Ω 
Number of inputs: one point 
DI function — One of the following functions can be 
assigned to the optional DI signal. 
Range switching — Selects either the higher or lower 
range in the unidirectional or bidirectional 2-range 
setting. 
Totalizer control — Starts and stops the built-in 
totalizer. 
Fixed-value outputs —Outputs fixed-values for 
current and pulse outputs. 
Zero adjustment — Executes zero adjustment 
(on-stream at zero flow rate). 
Output signals 
Current output: 
4–20mAdc (load resistance 0 to 750
Ω) 
Note: The current output cannot be used with the 
PROFIBUS-PA communication. 
Digital outputs — One point (std.) and one more 
point is optionally available as follows. 
Digital output DO1 (std.): 
Output type: Transistor open collector 
Number of outputs: One point 
Output capacity: 30Vdc, 200mA maximum 
Digital output DO2 (opt.): 
Output type:   
Solid-state relay output (non polarity) 
Number of outputs: One point 
Output capacity: 150Vdc, 150mA maximum   
or 150Vac (peak to peak), 100mA maximum 
DO1 and DO2 functions — One of the following 
functions can be assigned to DO1 (std.) and/or 
DO2 (opt.) 
• Pulse output (available only for DO1, DO2) 
Pulse rate: 3.6 to 36,000,000 pulses/hr (DO1) 
3.6 to 360,000 pulses/hr (DO2) 
(Over 3,600,000 pulses/hr, auto-setting) 
Pulse width: 0.5 to 500ms (but less than half of 
the period for 100% flow rate) 
Note: The same and simultaneous   pulse is not 
available between DO1 and DO2.) 
• Multi-range selection outputs (Note 1) 
• High, High high, Low, and/or Low low alarm 
outputs (Note 2) 
• Empty pipe alarm output 
• Digital Output Active Status (DO1 and DO2) 
(Note 2) 
• Preset count output 
• Converter failure alarm output 
Note 1: Two outputs (DO1 and DO2) are needed for 
4-range switching and forward/reverse 2-range 
switching. 
Note 2:Normal Open (default set) or Normal Close is 
selected for alarm outputs when programming. 
When power failure occurs, unit will be fault to 
Normal Open.