Oritron ICS-2100 用户手册

下载
页码 288
5 • Service
Doc. 065291-01 3/09
101
Use the Wellness Panel to perform the following functions:
Calibrate the pump flow rate
Calibrate the vacuum degas assembly
Calibrate the conductivity cell
Test the conductivity cell electronics, using a dummy cell
Upload calibration values from the ICS-2100 to the Chromeleon 
Wellness database
After completing a calibration procedure from the touch screen,
always update the Chromeleon Wellness database by clicking the
Upload button on the Chromeleon Wellness Panel (see 
This action ensures that the current calibration information is stored
in Chromeleon as well as in the ICS-2100.