Oritron ICS-2100 用户手册
5 • Service
Doc. 065291-01 3/09
101
Use the Wellness Panel to perform the following functions:
•
Calibrate the pump flow rate
•
Calibrate the vacuum degas assembly
•
Calibrate the conductivity cell
•
Test the conductivity cell electronics, using a dummy cell
•
Upload calibration values from the ICS-2100 to the Chromeleon
Wellness database
Wellness database
After completing a calibration procedure from the touch screen,
always update the Chromeleon Wellness database by clicking the
Upload button on the Chromeleon Wellness Panel (see
always update the Chromeleon Wellness database by clicking the
Upload button on the Chromeleon Wellness Panel (see
This action ensures that the current calibration information is stored
in Chromeleon as well as in the ICS-2100.