ICS-2100 Ion Chromatography System
106
Doc. 065291-01 3/09
5.1.4
Calibrating the Flow Rate
When to Calibrate
If you run the ICS-2100 Operational Qualification or Performance 
Qualification and it fails.
Items Needed
0.076-mm (0.003-in) ID yellow PEEK tubing (P/N 049715) to create 
14 ± 2 MPa (2000 ± 300 psi) of backpressure (if needed)
High purity ASTM filtered, Type I (18 megohm-cm) deionized water
Balance capable of weighing more than 10 g with 0.001 g readability 
Tared beaker
To Calibrate
1. Open the ICS-2100 Wellness Panel in Chromeleon (see 
), or open the ICS-2100 touch screen 
FLOW RATE
 page 
2. Verify that there is 14 ± 1.4 MPa (2000 ± 200 psi) of backpressure.
Figure 5-5.  Flow Rate Page