ICS-2100 Ion Chromatography System
130
Doc. 065291-01 3/09
5.11 Replacing the Suppressor
Refer to the suppressor manual for guidance about when to replace a suppressor. 
Suppressor manuals are included on the Dionex Reference Library CD-ROM 
(P/N 053891).
1. Refer to the suppressor quick start guide (shipped with the suppressor) for 
instructions on preparing the suppressor before initial use and for additional 
installation details.
2. Turn off the pump from the ICS-2100 Control panel in Chromeleon or from 
the touch screen 
HOME
 page.
3. Open the front door of the ICS-2100.
4. Disconnect the two eluent and the two regenerant lines from the suppressor.
5. To remove the suppressor from the component panel, slide it up to detach it 
from the mounting tabs on the panel and then pull it toward you.
6. Continue pulling the 
suppressor out until the 
yellow connector on the 
cable inside the ICS-2100 is 
outside the panel (see 
7. Disconnect the suppressor 
cable (twist the ring on the 
yellow connector and pull the 
two connectors apart).
Figure 5-21.  Removing the Suppressor
8. Connect the cable from the new suppressor to the ICS-2100 cable (align the 
pins on the two connectors and push them together).
9. Orient the suppressor with the 
REGEN OUT
 fitting on top and the cables to the 
right. 
Yellow
Connector