Oritron ICS-2100 用户手册

下载
页码 288
ICS-2100 Ion Chromatography System
210
Doc. 065291-01 3/09
Touching 
 returns you to the 
HOME
 page.
Touching 
 opens a list of options (see 
B.2
Using the Touch Screen with Chromeleon
When the ICS-2100 is connected to a Chromeleon Control 
panel, the status field at the bottom center of the 
HOME
 page 
displays 
Remote Mode
. In this mode, most of the controls on the touch screen 
pages are disabled. This prevents touch screen commands from conflicting with 
Chromeleon commands. 
In general, when the ICS-2100 is connected to Chromeleon, you can:
View all ICS-2100 operational status information (pressure, flow rate, 
conductivity readings, etc.)
Go to any of the touch screen pages
View a plot of the conductivity data from the 
PLOT
 page and adjust the plot 
display using the controls on the page. See 
To restore touch screen (local) control, clear the Connected check box on the 
Chromeleon Control panel.
Figure B-5.  Suppressor Page: Selecting an Option
Touch an 
option to 
select it.
Note: To cancel 
and retain the 
current selection,
touch outside of 
the list.