ICS-2100 Ion Chromatography System
214
Doc. 065291-01 3/09
B.4.2
Home Page EGC Controls
The 
EGC
 controls indicate the 
type and the eluent 
concentration of each installed 
EluGen Cartridge (EGC). The 
CR-TC
 field indicates whether 
the Continuously Regenerated 
Trap Column is on or off.
To determine the eluent concentration for your application, refer to the column 
manual. Column manuals are included on the Dionex Reference Library CD-
ROM (P/N 053891). When the ICS-2100 is controlled from the front panel, only 
isocratic eluent delivery is possible. With isocratic delivery, the eluent 
composition and concentration remain constant throughout the run. When the 
ICS-2100 is under Chromeleon control, gradient delivery, in which the eluent 
concentration changes over time, is also possible.
To set the eluent concentration or to turn on or off the CR-TC, touch the 
concentration button. The 
ELUENT GENERATOR 
page appears (see 
NOTE When the EGC-2 is configured as an auxiliary power
supply, the EGC-2 cartridge type field displays 
AUX
 and
the EGC-2 eluent concentration field displays the actual
current (in mA) of the auxiliary power supply.
B.4.3
Home Page Column Heater
The temperature of the column heater can be controlled to 
between 30 and 60 °C. For optimal performance, set the 
column heater to at least 5 °C above the ambient 
temperature. To set the temperature, touch the 
Column 
Heater
 field and enter the desired temperature on the number pad. To turn the 
column heater off, set the temperature to 0. The display to the right of the 
temperature setting indicates the heater status:
column temperature is at equilibrium
column temperature is rising
column temperature is falling
Off
column heater is off
EGC-1 Eluent 
Concentration
EGC-1
Cartridge
Type
EGC-2
Cartridge
Type
EGC-2 Eluent 
Concentration