B • Touch Screen Operation
Doc. 065291-01 3/09
229
below 30%. At 0%, an error message informs you that the ion count is 
depleted and the cartridge must be replaced before continuing 
operation.
NOTE When the EGC-2 power supply is configured as
an auxiliary power supply, the EGC-2 Life
Remaining field is blank.
Normalized Life Used
 (displayed when an EPM is installed) 
indicates how much the EPM has been used. The value starts at 0 and 
counts up with use. A count of 1 means the EPM usage is 
approximately equal to the lifetime of one K
2
CO
3
 cartridge. 
B.8.5
Controlling the CR-TC Power
To manually control the CR-TC power, touch the 
CR-TC Off
 and 
On
 
buttons on the 
ELUENT GENERATOR
To automatically control the CR-TC power simultaneously with the 
pump flow, select 
Auto turn on with pump
.
NOTE The CR-TC power always turns off when the pump
is off, regardless of whether the “Automatically turn
on with pump” option is selected.
Figure B-17.  Eluent Generator Page: CR-TC Controls