ICS-2100 Ion Chromatography System
20
Doc. 065291-01 3/09
Suppressor
The suppressor reduces the eluent conductivity and enhances the 
conductivity of the sample ions, thereby increasing detection sensitivity. 
Either an AES
®
 Atlas Electrolytic Suppressor, SRS
®
 Self-Regenerating 
Suppressor, or MMS
 MicroMembrane
 Suppressor can be used with 
the ICS-2100. See 
 for details about the suppressor.
Separator and Guard Columns
Both the separator and guard columns are packed with resin and perform 
the separation of the sample ions. The main function of the guard column 
is to trap contaminants and remove particulates that might damage the 
separator column.
Column Heater
The column heater controls the temperature of the separator and guard 
columns. The temperature can be set to between 30 and 60 °C; however, 
it must be set to at least 5 °C above the ambient temperature. See 
 for details about the column heater.
Eluent Valve
The eluent valve controls the flow from the eluent reservoir. The eluent 
valve opens automatically when the pump is started and closes when the 
pump is turned off. 
Screen Adjustment Knob
Use this knob to change the LCD touch screen contrast.
Tubing Chases
The upper tubing chase routes tubing from the top cover to the component 
panel. The lower tubing chase routes tubing from the component panel, 
through the interior of the ICS-2100, to the rear panel.