ICS-2100 Ion Chromatography System
54
Doc. 065291-01 3/09
3.4
Set Up the Eluent Reservoir
The ICS-2100 does not require pressurized reservoirs. However, if eluent is 
manually degassed or is sensitive to contamination, Dionex recommends 
pressurizing the reservoir with helium or nitrogen. 
The air regulator accessory bracket and other items needed for pressurizing the 
eluent reservoir must be ordered separately (P/N 060054). For more information, 
refer to the ICS-2100 installation instructions.
3.4.1
Filter the Deionized Water
Filtering removes small particulates in the deionized water that may 
contaminate the pump check valves and cause erratic flow rates or loss of 
prime. An end-line filter (P/N 045987) is provided in the ICS-2100 Ship 
Kit (P/N 064375) for this purpose. 
Install the end-line filter on the end of the deionized water line, inside the 
reservoir. Verify that the end of the filter extends to the bottom of the 
reservoir and that the filter is submerged in deionized water. This prevents 
air from being drawn through the lines.
3.4.2
Fill the Reservoir
Fill the reservoir with ASTM filtered, Type I (18-megohm) deionized 
water.
3.4.3
Set the Eluent Level
After filling the reservoir, enter the volume of liquid in the reservoir in 
one of the following locations:
On the 
HOME
 page in the 
Eluent Level
In the Eluent Bottle field in the Pump Settings window (accessed 
from the ICS-2100
 
Control panel on the Chromeleon panel tabset (see 
The ICS-2100 determines the eluent usage by monitoring the flow rate 
and the length of time the pump is on, and updates the Eluent Bottle 
volume as the eluent is depleted. A warning appears if the level falls 
below 200 mL. Warnings are repeated at 100 mL and 0 mL.