21.4
Performing meter validation
To perform meter validation, measure a sample of the process fluid and compare the measurement
with the flowmeter’s reported value.
with the flowmeter’s reported value.
Use the following formula to calculate a meter factor:
Valid values for meter factors range from
0.8
to
1.2
. If the calculated meter factor exceeds these
limits, contact Micro Motion customer service.
To enter the new meter factor, use the Inputs menu (see Figure 7-1).