Micro Motion
®
 Series 3000 MVD Transmitters and Controllers
Measurement Performance
Figure 21-9
Smart Meter Verification scheduler – Display
Figure 21-10
Smart Meter Verification scheduler – Communicator
21.4
Performing meter validation
To perform meter validation, measure a sample of the process fluid and compare the measurement 
with the flowmeter’s reported value.
Use the following formula to calculate a meter factor: 
Valid values for meter factors range from
 0.8
 to 
1.2
. If the calculated meter factor exceeds these 
limits, contact Micro Motion customer service.
To enter the new meter factor, use the Inputs menu (see Figure 7-1).