National Instruments NI PXI-5422 用户手册

下载
页码 63
© National Instruments Corporation
11
NI PXI-5422 Calibration Procedure
Figure 2.  Optional Calibration Programming Flow
Self-Calibrate
Verify
Document 
Pre-Adjustment
Results
Meets
Calibration
Test Limits?
Meets
Published
Specifications?
Update Calibration
Dates and
Temperatures
Without Adjusting
Results
Within
Published Specs
(Adjustment
Optional)
Adjust
Anyway?
Adjust
(Cal Dates and 
Temperatures
Updated)
Verify
Document 
Post-Adjustment
Results
Meets
Calibration
Test Limits?
Calibration/
Verification
Complete
Review
Verification/
Adjustment
Procedure or
Return Device
Yes
No
Yes
No
Yes
No
Yes
No