NI PXI-5422 Calibration Procedure
12
ni.com
External Calibration Requirements
This section describes the test equipment and test conditions required for 
calibration.
Test Equipment
External calibration requires different equipment for each applicable 
specification. Refer to Table 2 for a list of equipment.
Test Conditions
Follow these guidelines to optimize the connections and the environment 
during calibration:
Keep connections to the NI PXI-5422 short.
Keep relative humidity below 80%.
Maintain a temperature between 18 
°C and 28 °C.
Observe the 15-minute warm-up time.
Table 2.  Equipment Required for Calibrating the NI PXI-5422 
Instrument
Applicable 
Specification
Minimum 
Specifications
Recommended 
Instrument
Digital multimeter 
(DMM)
 DC gain and offset
DC accuracy 
≤ ±50 ppm
Resolution 
≤1 µV
NI PXI-4070
Agilent HP 34401A
Keithley 2000
Male banana to 
female BNC adapter
Male BNC to 
female SMB cable
50 
Ω, RG-223
Spectrum analyzer or 
frequency meter
Frequency accuracy
Ability to measure 
10 MHz or higher 
sine waves
Frequency accuracy 
to ±500 ppb
NI PXI-5660
Agilent HP 8560E
Agilent HP 53131A 
or HP 53132A 
with timebase option 
001, 010, or 012
Male BNC to 
female SMB cable
50 
Ω, RG-223