NI PXI-5422 Calibration Procedure
32
ni.com
11. Measure the negative DC voltage out of the NI PXI-5422. Record the 
negative DC voltage out measurement in the Measured Negative 
Full-Scale Value 
column of Table 6.
12. Call 
niFgen_AbortGeneration
 (niFgen Abort Generation VI) to 
abort the waveform generation using the following parameter:
vi: The session handle returned from 
niFgen_init
13. Average the Measured Positive Full-Scale Value and Measured 
Negative Full-Scale Value to calculate the Offset
14. Verify that the Offset is less than or equal to the Offset Limit listed in 
Table 6 for the current iteration. 
15. Subtract the Offset and the Ideal Full-Scale Value from the Measured 
Full-Scale Value to get the Error Full-Scale Value for both the positive 
and negative settings, respectively. 
16. Return to step 4 until all iterations are completed.
17. Verify that these errors are less than or equal to the limits you are using. 
If any of the errors are greater than the Calibration Test Limit or the 
Published Specification, perform an external adjustment.
18. Call 
niFgen_close
 (niFgen Close VI) to close the instrument driver 
session, to destroy the instrument driver session and all of its 
properties, and to release any memory resources that NI-FGEN 
uses. Use the following parameter:
vi: The session handle returned from 
niFgen_init
Adjusting the NI PXI-5422
If the NI PXI-5422 successfully passes all verification within the 
calibration test limits, adjustment is recommended, but not required, to 
guarantee its published specifications for the next two years. If the 
NI PXI-5422 was not within the calibration test limits for each verification 
procedure, perform the adjustment procedure to improve the accuracy of 
the NI PXI-5422. Refer to the 
 section to 
determine which procedures to perform.
An adjustment is required only once every two years. The adjustment 
procedure automatically updates the calibration date and temperature in 
the EEPROM of the NI PXI-5422. 
If the NI PXI-5422 passed verification within the calibration test limits and 
you do not want to do an adjustment, you can update the calibration date 
and onboard calibration temperature without making any adjustments by 
completing the following steps: