NI PXI-5422 Calibration Procedure
44
ni.com
g.
Call 
niFgen_WriteBinary16AnalogStaticValue
 (niFgen 
Write Binary 16 Analog Static Value VI) to set the main DAC 
value using the following parameters:
vi: The session handle returned from 
niFgen_InitExtCal
 
channelName
"0"
value: –
29232
h.
Call 
niFgen_Commit
 (niFgen Commit VI) to commit the 
attribute values to the device using the following parameter:
vi: The session handle returned from 
niFgen_InitExtCal
i.
Wait 500 ms for the output to settle.
j.
Use the DMM to measure the voltage output by the device. This 
measurement is 
measurement 1
8.
Call 
niFgen_CalAdjustMainPathPreAmpGain
 (niFgen Cal 
Adjust Main Path Pre Amp Gain VI) using the following parameters:
vi: The session handle returned from 
niFgen_InitExtCal
channelName
"0"
configuration: The Current Configuration value for the current 
iteration from Table 8
mainDACValues: An array containing two elements—the two 
values (
25233
, –
29232
) that you set for the main DAC in the 
order that you measured them
gainDACValues: An array containing two elements—the two 
values (
1500
2000
) that you set for the gain DAC in the order 
that you measured them
offsetDACValues: An array containing one element—the value 
(
32000
) that you set for the offset DAC
measuredOutputs: An array containing two elements—the 
two output voltages (
measurement 0
measurement 1
) that 
you measured in the order that you measured them