National Instruments NI PXI-5422 用户手册
NI PXI-5422 Calibration Procedure
44
ni.com
g.
Call
niFgen_WriteBinary16AnalogStaticValue
(niFgen
Write Binary 16 Analog Static Value VI) to set the main DAC
value using the following parameters:
value using the following parameters:
•
vi: The session handle returned from
niFgen_InitExtCal
•
channelName:
"0"
•
value: –
29232
h.
Call
niFgen_Commit
(niFgen Commit VI) to commit the
attribute values to the device using the following parameter:
•
vi: The session handle returned from
niFgen_InitExtCal
i.
Wait 500 ms for the output to settle.
j.
Use the DMM to measure the voltage output by the device. This
measurement is
measurement is
measurement 1
8.
Call
niFgen_CalAdjustMainPathPreAmpGain
(niFgen Cal
Adjust Main Path Pre Amp Gain VI) using the following parameters:
•
vi: The session handle returned from
niFgen_InitExtCal
•
channelName:
"0"
•
configuration: The Current Configuration value for the current
iteration from Table 8
iteration from Table 8
•
mainDACValues: An array containing two elements—the two
values (
values (
25233
, –
29232
) that you set for the main DAC in the
order that you measured them
•
gainDACValues: An array containing two elements—the two
values (
values (
1500
,
2000
) that you set for the gain DAC in the order
that you measured them
•
offsetDACValues: An array containing one element—the value
(
(
32000
) that you set for the offset DAC
•
measuredOutputs: An array containing two elements—the
two output voltages (
two output voltages (
measurement 0
,
measurement 1
) that
you measured in the order that you measured them