National Instruments NI PXI-5422 用户手册

下载
页码 63
NI PXI-5422 Calibration Procedure
46
ni.com
6.
Call 
niFgen_SetAttributeViInt32
 to set the analog path value 
(niFgen property node: Output Attributes»Analog Path) using 
the following parameters:
vi: The session handle returned from 
niFgen_InitExtCal
channelName
"0"
attributeID
NIFGEN_ATTR_ANALOG_PATH
value: The Analog Path value for the current iteration from 
Table 9
7.
Call 
niFgen_SetAttributeViReal64
 to set the post-amplifier 
attenuation (niFgen property node: Calibration»Post-Amplifier 
Attenuation
) using the following parameters:
vi: The session handle returned from 
niFgen_InitExtCal
channelName
"0"
attributeID
NIFGEN_ATTR_POST_AMPLIFIER_ATTENUATION
value: The Post-Amplifier Attenuation value for the current 
iteration from Table 9
Table 9.  Attributes and Values for the Main Path Post-Amplifier Gain and Offset 
Iteration
Analog Path
Post-Amp 
Attenuation
Current Configuration
1
NIFGEN_VAL_FIXED_LOW_GAIN_
 
ANALOG_PATH
0
NIFGEN_VAL_CAL_CONFIG_MAIN_PATH_
 
LOW_GAIN_0DB
2
NIFGEN_VAL_FIXED_LOW_GAIN_
 
ANALOG_PATH
12
NIFGEN_VAL_CAL_CONFIG_MAIN_PATH_
 
LOW_GAIN_12DB
3
NIFGEN_VAL_FIXED_LOW_GAIN_
 
ANALOG_PATH
24
NIFGEN_VAL_CAL_CONFIG_MAIN_PATH_
 
LOW_GAIN_24DB
4
NIFGEN_VAL_FIXED_LOW_GAIN_
 
ANALOG_PATH
36
NIFGEN_VAL_CAL_CONFIG_MAIN_PATH_
 
LOW_GAIN_36DB
5
NIFGEN_VAL_FIXED_HIGH_GAIN_
 
ANALOG_PATH
0
NIFGEN_VAL_CAL_CONFIG_MAIN_PATH_
 
HIGH_GAIN_0DB
6
NIFGEN_VAL_FIXED_HIGH_GAIN_
 
ANALOG_PATH
12
NIFGEN_VAL_CAL_CONFIG_MAIN_PATH_
 
HIGH_GAIN_12DB
7
NIFGEN_VAL_FIXED_HIGH_GAIN_
 
ANALOG_PATH
24
NIFGEN_VAL_CAL_CONFIG_MAIN_PATH_
 
HIGH_GAIN_24DB
8
NIFGEN_VAL_FIXED_HIGH_GAIN_
 
ANALOG_PATH
36
NIFGEN_VAL_CAL_CONFIG_MAIN_PATH_
 
HIGH_GAIN_36DB