Omega Engineering OMB-1108-0901 用户手册

下载
页码 50
DaqTemp Series User’s Manual                                                       
08-27-02
CE-Compliance     5-1
CE-Compliance
5
Overview
CE-compliant products bear the “CE” mark and include a Declaration of Conformity stating the
particular specifications and conditions that apply.  The test records and supporting documentation
that validate the compliance are kept on file at the factory.
The European Union established CE standards in 1985.  The standards include specifications for safety,
EMI emissions, and immunity from electromagnetic interference.  Products that are intended for placement
in the European Union must meet or exceed the standards and bear the "CE" mark.
Although not required in the USA, meeting or exceeding the CE standards is considered good engineering
practice, since doing so enhances safety while reducing noise and ESD problems.
In contracted and in-house testing, most Daq products met the required specifications.  In many cases,
products that were not originally in compliance were redesigned accordingly.  In noted instances, alternate
product versions, shield plates, edge guards, special connectors, or add-on kits are required to meet CE
compliance.
CE Standards and Directives
The electromagnetic compatibility (EMC) directives specify two basic requirements:
1.  The device must not interfere with radio or telecommunications.
2.  The device must be immune from electromagnetic interference from RF transmitters, etc.
The standards are published in the Official Journal of European Union under direction of CENELEC
(European Committee for Electrotechnical Standardization).  The specific standards relevant to Daq
equipment are listed on the product’s Declaration of Conformity and include: CISPR22:1985;
EN55022:1988 (Information Technology Equipment, Class A for commercial/industrial use); and
EN50082-1:1992 for various categories of EMI immunity.
The safety standard that applies to Daq products is EN 61010-1 : 1993 (Safety Requirements for Electrical
Equipment for Measurement, Control, and Laboratory Use, Part 1: General Requirements
). 
Environmental conditions include the following:
•  indoor use
•  altitude up to 2000 m
•  temperature 5°C to 40°C (41°F to 104°F)
•  maximum relative humidity 80% for temperatures up to 31°C (87.8°F) decreasing linearly
to 50% relative humidity at 40°C (104°F)
•  mains supply voltage fluctuations not to exceed ±10% of the nominal voltage
•  other supply voltage fluctuations as stated by the manufacturer
•  transient overvoltage according to installation categories (overvoltage categories) I, II and III
For mains supply, the minimum and normal category is II
•  pollution degree I or II in accordance with IEC 664