Philips V24CT 用户手册

下载
页码 392
oxyCRG Recordings
6-36     Recording Functions  
Re
c
o
rd
ing
 Func
tions
  
oxyCRG Recordings
You can record the contents of the oxyCRG display on an M1116B Plug-
In recorder.
Note—
oxyCRG recordings can only be made on the M1116B Plug-In 
recorder. The M1116A Plug-In recorder, the M1117A four channel 
recorder, and the central recorder do not support oxyCRG recording. 
You can manually request an oxyCRG recording, and you can configure 
an alarm recording of oxyCRG.
The oxyCRG recording is always 6 minutes prior to the most recent data 
on the screen. This enables you to have a hardcopy of events leading up 
to the current situation. When you stop recording, the recorder stops 
immediately, with the result that the most recent 6 minutes of patient 
data will not be recorded.
Note—
OxyCRG alarm triggered recordings differ slightly from the 
oxyCRG episode recordings generated by Neonatal Events. All standard 
oxyCRG recordings are realtime recordings, i.e. when the alarm occurs. 
An oxyCRG episode recording is initiated by the user to document 
selected events retrospectively.
Manual request:
An oxyCRG recording with a fixed delay of 
six minutes (corresponding to the display 
content) is produced. 
Alarm recording:
An automatic oxyCRG recording of six 
minutes of pre-alarm data and two minutes 
of post-alarm data is produced.