oxyCRG Recordings
Recording Functions   6-37
Re
c
o
rd
ing
 Func
tions
  
The oxyCRG is shown in three distinct non-overlapping “channels” on the 
strip. The time, the three wave labels, and the scales for the oxygen and 
beat-to-beat Heart Rate channels are printed periodically on the 
recording.
Three speeds are available for oxyCRG recording: 1, 2, or 3 cm/min. You 
can choose the speed in the oxyCRG Task Window
.
Note—
oxyCRG recordings made at slow speeds (1 or 2 cm/min) can cause 
deposits of paper debris on the print head over time. In this case, 
recordings may appear unevenly faint. To correct, see “Cleaning the Print 
head in the Plug-In Recorder” on page 6-52. Th
en reload the paper and 
continue recording. 
(8AMIN)                      2  cm/min                                                          05  JUL  9 11:20
05 JUL 94 11:20
HR 127
PULSE 125 ** PULSE 125 > 120
ABP 80/40 (59)
PAP -?- PAP  ZERO+CHECK CAL
SpO2 87  **SpO2  87 < 90
NBP 82/41 (60) 05 JUL 94 10:56
AWRR 40
T1 35.2
T2 -?- T2
SpO2  80
100
60
HR  140
180
100
RESP
Current
Patient Name
Date/time
recording was initiated
recorder speed
Numerics & Alarms
values at this point
were taken at
THOMAS SMITH 123789