Philips V24CT 用户手册

下载
页码 392
Types of Transfer
Data Transfer   10-13
D
a
ta
 T
ra
n
sf
er
  
Transferring data from a source monitor via a transit monitor to 
a destination monitor requires special handling if:
The source monitor and the destination monitor are:
— CMS Release G or later or
— V24 or V26 Release A.0 or later
and
The transit monitor is:
— CMS Release E or F
— V24 Release E
— CTS
1. Plug the Data Transfer Module into the rack of the source 
monitor.
2. Press the 
 hardkey on the module and transfer all 
data to the module.
3. Plug the Data Transfer Module into the rack of the transit 
monitor.
4. Select 
 
in the Data to Module task window 
to collect new data to the module without clearing the module.
5. Remove the module from the transit monitor and plug it into 
the destination monitor.
6. Press the 
 hardkey on the module and transfer the 
data to the monitor.
Note—
With specific configurations, data loss can occur when transferring 
data to a V24 or V26 Release B.0 via the Data Transfer module.
Module
Collect New Data
Pts
Refernce
Monitor