Combining Data
10-14     Data Transfer  
D
a
ta
 T
ra
n
sf
er
  
Combining Data
In some situations, the patient name and/or ID in the monitor and 
module may not match (see Append New Data display on previous 
page). In other cases, one or both names may be blank. The Data 
Transfer Combine Task Window enables you to combine files when the 
demographic data is missing or different.
Time 
Conversion
Data in a monitor can be stored at either 1-minute or 12-second 
resolution, depending on the monitor's configuration. Data in a module 
is stored only at 1-minute resolution. Time conversions are done 
automatically when the resolutions are different.
When converting data from 12-second to 1-minute resolution (that is, 
from a monitor to a module), one data sample out of 5 is selected to 
represent the minute. The sample selected is the one on the minute 
boundary.
When transferring data from 1-minute to 12-second resolution (from a 
module to a monitor), one valid sample and 4 blanks will be used for 
each minute. The valid sample will be placed on the minute boundary.
Database 
Conversion 
Transfers to the Module
A module contains enough memory to store 24 hours of data for 16 
parameters at 1 minute resolution. If the data to be transferred does not 
fit in the module, then the number of hours will be reduced to fit the 
available memory.
The following table shows the conversion between the monitor and 
module databases during a transfer to the module.