AC and DC (Battery) Operation
12-2     Battery Information (V24CT and V26CT only)  
Ba
tte
ry
 I
n
for
m
a
ti
on 
(V
24
CT
 an
d
 V
26C
T
 
AC and DC (Battery) Operation
You should only use Philips 40488A 12 Volt sealed lead-acid batteries 
obtained from Philips Medical Systems for use with the Philips M1205A 
V24CT. Use of other camcorder-type batteries could result in inaccurate 
estimates of battery capacity remaining or in shorter battery life.
Warning
Warning
When connecting a non-medical device such as a printer to the 
RS-232 Interface of V24CT, the monitor must be properly 
grounded, or the non-medical device properly isolated. This must 
be done in order to prevent the possibility of an electrical shock 
hazard (to the user and or patient) from excessive ground 
leakage currents from the non-medical device such as a printer, 
and to maintain compliance with IEC 601-1-1 where applicable. 
This hazard is especially prevalent when the V24CT is operated 
in battery mode.   
In normal AC line operation grounding the monitor is 
accomplished by ensuring that the AC line cord is connected 
from the monitor to a properly grounded AC receptacle. 
Alternatively and additionally, proper grounding may be achieved 
by attaching a safety grounded wire to the threaded safety 
ground stud 
 found on the rear of the monitor.
Isolation may be achieved by powering the printer through a 
safety isolation transformer.