AC and DC (Battery) Operation
12-4     Battery Information (V24CT and V26CT only)  
Ba
tte
ry
 I
n
for
m
a
ti
on 
(V
24
CT
 an
d
 V
26C
T
 
Note—
When AC is connected and the monitor is on, the Battery Charge 
LEDs may take some time to cycle to the appropriate charge indication. 
The V24CT or V26CT may under report battery capacity during this 
setting period. Use the fuel gauge rather than the Battery Charge LEDs 
during this period to estimate battery capacity or turn the monitor off to 
accelerate the charge setting time.
Battery LED Pattern
A fully depleted battery will charge to 90% of full capacity in 4 
hours when the monitor is not being used. When the monitor is in 
use, it will take approximately 16 hours to charge the batteries to 
90%.
To preserve battery integrity, the battery must be fully charged 
each time the battery is depleted. 
A new battery or one that has been stored for an extended period 
requires 4 hours of charging before use. 
To operate on internal battery power only, disconnect the power 
cord from the AC outlet. 
Batteries should be discarded if there are visual signs of damage
Life expectancy of a battery depends on the frequency and 
duration of use. When properly maintained and stored, the life 
expectancy of a battery is about 1.5 years. For more aggressive use 
models, life-expectancy may be less. Philips recommends 
replacing sealed lead acid batteries once a year. The date of 
manufacture can be found on the battery. It is coded as yymmdd. 
Residual Capacity
Charging LED
Charged LED
Up to 30-40%
Flashing
Off
More than 30-40%, Less then 90% Steady
Off 
More than 90%
Off
Steady
Full
Off
Steady