Performance Assurance Checks
Maintenance   13-15
M
a
int
e
na
nc
e
  
Performing 
the ECG 
Module and 
ECG/RESP 
Self-Test
Auto Check
 - None
Self-Test
 The ECG and ECG/RESP self-test may be performed at any 
time as the lead selector and pace pulse detection are inhibited during the 
check. The module will produce ECG test signals at the beginning of the 
processing circuitry, and process the test signals in the same way patient 
signals are processed. The test signals last about 30 seconds and then the 
display returns to the normal monitoring mode.
TEST - Press 
 
 
.
RESULT - If the module is functioning correctly, the following heart rate 
and respiration are displayed without the alarms being activated:
ECG 100 bpm in ADULT mode 125 bpm in NEO/PEDI mode
RESP 15 rpm in ADULT mode 30 rpm in PEDI mode 55 rpm in NEO mode
Cardiac Output
5 l/min (+0.1 l/min) 
(Blood Temp: 37°C) 
(Inj Temp: 0°C)
No wave
tcpO
2
/tcpCO
2
No wave 
tcpO
2
60 mmHg (8.0 kPa)
tcpCO
2
40 mmHg (5.3 kPa)
CO
2
Simulated square wave 
ETCO
2
40 mmHg (6.0 kPa)
IMCO
2
0 mmHg (0.0 kPa)
AWRR
25 rpm
Temperature
40°C (104°F)
No wave 
Module 
Test Numeric 
Test Waveform
Monitor Setup
Test Signals
Pts
Refernce
Signals