Performance Assurance Checks
Maintenance   13-17
M
a
int
e
na
nc
e
  
Performing 
the NBP 
Module Self-
Test
Auto Check
 - The NBP module performs a check periodically during 
operation.
Self-Test
 - None
Performing 
the SpO
2
/
Pleth 
Module Self-
Test
Auto Check
 - None
Self-Test - The SpO
2
/Pleth module self-test may be performed at any time. 
The integral waveform generator in the module produces test signals 
using the current driver at the beginning of the processing circuitry. The 
test signals are processed in the same way as patient signals are 
processed. The test signals last about 30 seconds and then the display 
returns to the normal monitoring mode.
TEST - Press 
 
 
.
RESULT - If the module is functioning correctly, the following Derived 
values and waveforms are displayed without the alarms being activated:
SpO
2
 100%
Pleth Wave height after auto gain is 100% of scale height
Pleth Pulse Rate 100 bpm in ADULT mode 125 bpm in NEO/PEDI mode
Performing 
the Cardiac 
Output 
Module Self-
Test
Auto Check
 - None
Self-Test
 - The Cardiac Output module test may be performed at any 
time. The test signal lasts about 30 seconds and then the display returns 
to the normal monitoring mode.
TEST - Press 
 
 
RESULT - If the module is functioning correctly, the following derived 
values will be displayed without the alarms being activated:
Cardiac Output 5.0 l/min +0.1 l/min
TBlood 37C + 0.1C
Tinj 0C + 0.1C
Monitor Setup
Test Signals
Pts
Refernce
Signals
Monitor Setup
Test Signals
Pts
Refernce
Signals