Philips M1500T 用户手册

下载
页码 27
  
10 
3
2
1
3
1 2
TOP VIEW
FRONT VIEW
PHOENIX CONTACT P.N.  1777992
Connecting Power 
The M1500 monitor is powered from 100-240 VAC, 50/60 Hz or optionally from 24 VDC. Damage 
will occur if 100-240 VAC power is connected to an M1500 equipped with the 24 VDC input power 
option.
 
M1500’s equipped with the 24 VDC option will have a “-24” suffix in their model number such as 
M1500-24 or M1500T-24. 
Because the M1500 is UL 1604 listed for Hazardous Location use, (Class I Division2, Groups A, 
B, C, D; Class II Division 2 Groups F and G: Temperature Code T4A), it has no power switch for switching 
off  supplied  power.  Please  read  the  functional  description  of  the  POWER  button  on  page  12. 
Consideration  should  be  give  to  the  installation  of  an  appropriately  rated  external  power  switch  if  the 
application requires powering off the M1500. 
Power  is  connected  to  the  M1500  through  a  removable  Phoenix  Contact  plug  (Phoenix  Contact 
P.N. 1777992) that allows for screw termination  of field wiring. The use of 18 AWG or greater (12 AWG 
maximum) is recommended. Connect the field  wiring  according to the appropriate table below. After the 
connections  are  made, make  sure  the 
plug retention screws (the two screws shown in the “Front View” 
below) are securely tightened. This will prevent the plug from pulling out. 
 
 
 
 
 
 
 
 
 
 
 
 
 
100 VAC 
– 240 VAC INPUT (1.0 Amps Min) 
 
PIN No. 
Definition 
AC Line Input 
AC Neutral Return 
Protective Earth Ground 
 
18 VDC 
– 36 VDC INPUT (2 Amps Recommended) 
 
PIN No. 
Definition 
+DC Input 
-DC Return 
Protective Earth Ground 
3
1 2
3