Performing and Reviewing Calculations
Trends and Calculations   8-21
Tr
e
nds
 a
nd 
C
a
lcula
ti
o
n
s
  
Performing 
Calculations
Calculations can be performed using automatically-entered values or 
manually-entered values. If a parameter is being continuously monitored, 
its value is automatically entered. Values which are not monitored must 
be entered manually. You can also change any of these values if needed.
In hemodynamic calculations, if the systolic and diastolic pressures are 
manually entered via the cursor keys or the handheld keypad, the mean 
pressure is calculated, marked with an asterisk, and stored automatically. 
The following formula will be used to estimate the mean pressure:
systolic + (diastolic x 2) / 3
Both indexed and non-indexed values are calculated. Calculated and input 
values are first presented with their associated units. If desired, the units 
column for calculated values may be replaced with normal ranges, and 
the input units column may be replaced with the times at which the 
values were stored in the database. 
Parameters that are sourced in kPa, cmH
2
O or mbar are automatically 
converted into mmHg to provide consistent indexed and non-indexed 
calculation units. Ventilation calculations convert input pressures from 
mmHg, kPa, and mbar into cmH
2
O before calculating output values.